Versions Compared

Key

  • This line was added.
  • This line was removed.
  • Formatting was changed.
Info

TAs: Süleyman Yasin Peker

Topics covered:

  1. Introduction to Verasonics Vantage 32LE Ultrasound Research System.

  2. Introduction to ultrasound transducers.

  3. Introduction to ultrasound imaging modes.

  4. Real-time data acquisition using Verasonics.

  5. Time-of-flight Flight measurement of common carotid artery (cca) using ultrasound and image processing.

  6. Flow measurement and colour doppler imaging using ultrasound.

Experiment details:

Introduction

This experiment involves the use of the Verasonics Vantage 32LE, a sophisticated ultrasound research system, to explore various aspects of ultrasound imaging and data acquisition. Participants will gain hands-on experience with different ultrasound imaging modes, including A-Mode, B-Mode, Colour Doppler, and Power Doppler, while understanding the applications of linear array, curvilinear/convex, and phased array transducers. The experiment will also cover real-time data acquisition using the Verasonics system, focusing on measuring timeTime-of-flight Flight values for echoes and analyzing the power spectrum of ultrasound signals in Doppler mode. This comprehensive approach will provide a deeper understanding of both the technical and practical aspects of ultrasound technology.

...

The earliest ultrasound mode showed returning echoes in a one dimensional, graphical format (Figure 3). Known as amplitude or A-mode, the information could be used to measure the distance between, or thickness of, tissues. An example is examining the cornea, lens and chambers of the eye.  

“B” or Brightness mode represents the amplitude peaks seen in A-mode as dots or pixels of varying brightness (Figure 2). 

...

Using B-mode, ultrasound systems can send sequential ultrasound pulses out in different directions to form multiple image lines. This process is completed quickly and repeatedly, creating the typical ultrasound image seen on all systems.

In this display mode, Doppler data received by the ultrasound system is displayed as coloured pixels within the sample box shown on the B-mode image (Figure 6). This gives information about direction and a semi-quantitative assessment of blood flow velocities.

...

A variation of colour Doppler available on most systems is a power Doppler display. This mode ignores the directional information provided by Doppler shift and displays the total Doppler signal strength as shades of one colour (Figure 7). Whilst it does not display any data on flow direction, power Doppler is a useful tool for examining low velocity blood flow and is more sensitive to flow than colour doppler.

...

Time-of-Flight Measurement

Time-of-Fight (TOF) in ultrasound refers to the time it takes for an ultrasound pulse to travel from the transducer to a target (such as tissue or a vessel wall) and back to the transducer after being reflected. It is a critical measurement used to determine the depth of structures within the body, as the distance traveled by the ultrasound wave is directly related to the time it takes for the wave to return to the transducer.

...

The formula to calculate the time-of-flight for a single echo is:

...

where:

  • d is the depth of the target (the distance from the transducer to the reflecting surface),

  • c is the speed of sound in the medium (typically c≈1540 m/s in soft tissue),

  • The factor of 2 accounts for the round-trip distance (to the target and back).

...

Calculating Time-of-Flight for Two Echoes

When considering two echoes, the TOF for each can be calculated separately if the depths d1d_1d1​ and d2d_2d2​ are known:

...

To find the difference in time-of-flight between the two echoes:

...

Report requirements:

The report for this part is expected to include: 

...